Infantile 800 Incubator

The most censorious equipment for the cutest soul in the world

Infantile 800 Incubator

HEATING SOURCE DOUBLE PROTECTION

Infantile is equipped with both thermocouple and mechanical temperature switch to protect the heating source of heating system and evaporator. Thermocouple reads real time temperature and limits the temperature in a safe range by feedback the temperature value to control module. While mechanical temperature switch is in series circuits, if the heating source temperature goes unmoral level, temperature cuts off the circuit physically.

THREE CHANNEL TEMPERATURE SENSORS

the heating source temperature goes unmoral level, temperature switch cuts off the circuit physically. There are three channel of temperature sensors in the module box. One channel is for reading hood temperature to remain the temperature at setting value. Another temperature sensor is in an individual circuit for protection only. The third sensor is used to compare the reading with other two sensors. If the reading of three sensors differ from 0.8°C, alarm is triggered.

TEMPERATURE UNIFORMITY

Infantile incubator's air circuts system is designed based on massive experimentation. With plenty aerodynamics knowledge, the temperature uniformity is improved as much as possible in order to give infant the safest micro-environment.

STABLE

AMBIENT TEMPERATURE MONITORING

With ambient temperature probe Infantile incubator is able to compare the gap between ambient and setting temperature and changing heating power output accordingly in order to provide more stable inside hood air temperature.

HEATING SYSTEM MONITORING

Infantile monitors not only hood air temperature but also radiator and circuit air temperature. By collecting each parts' temperature information, Ecol-a's temperature control system is much more accurate and safe than others.

PID FUZZY ALGORITHM

With the help of latest PID technology, Infantile es the temperature variability to unprecedented 0.2 C and temperature overshoot is also reduced dramatically.

BACKUP BATTERY

Infantile equipped with backup battery, once power supply is off, the backup battery supports the air circuit and alarm system keep working at least 30 minutes.

ALARM LIST

Infantile provides wide range of physiological and technical alarms in order to improve the treatment effect and eliminate residual risk from the device.

System Fault Alarm
Stuck Key
This alarm occurs when a stuck key is detected.
Disconnect
This alarm occurs when communication with the sensor module fails.
Sensor Module Failure3
This alarm occurs when the controller detects a sensor module fan is not rotating.
Sensor Module Failure6
This alarm occurs when the controller detects the ambient temperature sensor is open or shorted.
Sensor Module Failures
This alarm occurs when the controller detects that the sensor module watchdog reset.
Low Air Flow
This alarm occurs when an air circulation failure is detected.
Air Flow Probe Failed
This alarm occurs if the airflow probe connection is open or short circuited.
Motor Failed
This alarm is activated when the fan motor speed falls outside specified tolerances.
Power Failure
This alarm occurs when the AC power supply is off(within 30s)
Battery Disconnect
This Alarm occurs when battery is disconnected.
Sensor Out of Position
This alarm occurs when the sensor module is not in the hood.
Access Panel Open
This alarm occurs when access panel is open.
Heter Failed1
This alarm occurs when the heater is over temperature.
Heater Failed2
This alarm occurs when the heater is damaged.
Air Probe Failed
This alarm occurs when one of the two thermistors in the sensor module differ from the other by 0.80C or both two thermistors have a measurement error exceeding acceptable limits.
Skin1 Probe Fail Alarm
This alarm occurs in the Skin mode if the Skin1 probe is electrically open or shorted.
Watchdog Failed
This alarm occurs when a failure is detected with the watchdog.
Humidity Alarm
Humidity Heater Failed 1
This alarm occurs when the humidity heater temperature is over heat.
Add Water
NO enough water in water reservoir.
LOW Humidity
This alarm occurs when the humidity is 10% lower than setting humidity. Note: The alarm is blocked for 30 minutes everytime turn on the machine, or 15 minutes everytime the humidity setting value is changed.
Reservoir Out of Position
This alarm occurs when the reservoir not in position.
Humidity Heater Failed 2
This alarm occurs when the heater thermocouple wires are open or shorted.
Temperature Alarm
High Skin Temperature
This alarm occurs when the indicated displayed temperature differs from the Set temperature by > 0.5°C or > 1.0°C (user selectable, default 1.0
Low Skin Temperature
This alarm occurs when the indicated displayed temperature differs from the Set temperature by < 0.5°C or < 1.0°C (user selectable, default 1.0
Remove Skin2 Probe
This alarm occurs when two skin probes are installed and the Skin mode is selected.
Skin Probe Disconnect
This alarm is activated when the Skin1 temperature probe (only in the Skin mock) is removed from the Sensor module. The associated monitoring display is blanked. (within 30s)
High Temp CutOut
Under air control, this alarm is activated if the displayed incubator temperature reaches 38°C for set temperatures 37°C. Under skin control, this alarm is activated if the incubator temperature reaches 40°C for any set temperature.
High Air Temperature
This alarm occurs when the indicated displayed temperature differs from the set temperature by >1.5°C.
Low Air Temperature
This alarm occurs when the indicated displayed temperature differs from the set temperature by 2.5°C.
High Skin1 Temperature
This alarm occurs when the Air mode is enabled and the infant skin temperature (from the Skin 1 probe) is >38.0°C when the Override mode is not active, or when the >37°C mode Of operation is active.
High Skin 2 Temperature
This alarm occurs when the Air mode is enabled and the infant skin temperature (from the Skin2 probe) is >38.0°C when the Override mode is not active, or >39.0°C when the >37°C mode Of operation is active.
Oxygen Alarm
Oxygen Cell
This alarm occurs when the oxygen cell readings differ by more than 3%. As a result, the oxygen flow into the system is interrupted.
Difference
This alarm occurs when the displayed oxygen value is >3% beow the oxygen set point.
Low Oxygen
This alarm when the displayed Oxygen value is >3% above the oxygen set point.
High Oxygen
This Alarm occurs when the oxygen cells are out of date.
Change Oxygen Cells
This Alarm occurs when the 02 source pressure is low or high
Check 02 supply
This Alarm occurs when proportional value is failed.
Servo 02 System Fail
This Alarm occurs when the oxygen cell is disconnected with module box.

COMFORTABLE

Infantile provides newborn a supper quiet environment. With the help of aerodynamics, Infantile's hood sound is reduced to 45dB. With great detail design, Infantile also devoted to provide medical personnel pleasure operation experience.